--- Page 1 --- Implementation, Characterization, and Evaluation of an Inexpensive Low-Power Low-Noise Infrasound Sensor Based on a Micromachined Differential Pressure Transducer and a Mechanical Filter OMAR MARCILLO AND JEFFREY B. JOHNSON Department of Earth and Environmental Science, New Mexico Institute of Mining and Technology, Socorro, New Mexico DARREN HART Sandia National Laboratories, Albuquerque, New Mexico (Manuscript received 14 June 2011, in final form 31 January 2012) ABSTRACT The implementation, characterization, and evaluation of a low-cost infrasound sensor developed at the Infrasound Laboratory at the New Mexico Institute of Mining and Technology (Infra-NMT) are described. This sensor is based on a commercial micromachined piezoresistive differential pressure transducer that uses a mechanical high-pass filter to reject low-frequency outband energy. The sensor features a low-noise, 2.02-mPa rms (0.5–2 Hz), 5.47-mPa rms (0.1–20 Hz), or 5.62-mPa rms (0.05–20 Hz), flat response between 0.01 and at least 40 Hz; inband sensitivity of 45.13 6 0.23 mV Pa21; and a nominal linear range from 2124.5 to 1124.5 Pa. Intended for outdoor applications, the influence of thermal changes in the sensor’s response has been minimized by using a thermal compensated pressure transducer powered by an ultralow drift (,5 ppm 8C21) and noise (,4mV from peak to peak) voltage reference. The sensor consumes a minimum of 24 mW and operates with voltages above 8 V while drawing 3 mA of current. The Infra-NMT specifications described above were independently verified using the infrasound test chamber at the Sandia National Laboratories’ (SNL’s) Facility for Acceptance, Calibration, and Testing (FACT), and the following pro- cedures are for comparison calibration against traceable reference stands in voltage and pressure. Because of the intended broad frequency response of this sensor, the testing chamber was configured in a double- reference sensor scheme. A well-characterized microbarometer (with a flat-amplitude response between 0.01 and 8 Hz) and a microphone (with a flat-amplitude response above 8 Hz) were used simultaneously in this double-reference test configuration. 1. Introduction Infrasound refers to atmospheric elastic waves prop- agating with frequencies between 0.02 and 20 Hz. Infrasound can be produced by natural (e.g., volcanic eruptions, earthquakes, and meteorites) and anthro- pogenic events (e.g., nuclear or chemical explosions). Because it experiences low atmospheric attenuation, infrasound can travel long distances and provide an ideal remote sensing tool to study source phenomenology and atmospheric wave propagation. For example, in volcanic environments, mass flux, gas velocity, and precise source location have been estimated from infrasonic records (Johnson et al. 2003; Vergniolle et al. 2004). In atmo- spheric studies, infrasound is one method applied to probe the atmosphere by estimating thermal gradients and wind fields (Antier et al. 2007; Le Pichon et al. 2005; Marcillo and Johnson 2010). The characteristics of infrasound recorded in a specific location depend upon the type and location of the source and the conditions of the intervening atmosphere. The magnitude of infrasonic signals from geophysical events, measured in units of pressure, typically range from millipascals (mPa) to several thousands of pascals (Pa; Arrowsmith et al. 2010). Infrasound sensors require high resolution and a wide dynamic range (DR) to detect within the infrasonic field. Infrasound sensors are also Corresponding author address: Omar Marcillo, Department of Earth and Environmental Science, MSEC 208 801 Leroy Place, Socorro, NM 87801. E-mail: omarcill@ees.nmt.edu SEPTEMBER 2012 M A R C I L L O E T A L . 1275 DOI: 10.1175/JTECH-D-11-00101.1  2012 American Meteorological Society Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 2 --- required to reject outband energy, such as meteorolog- ical pressure fluctuations (very low frequency, ,0.01 Hz) and wind and anthropogenic noise (high frequency, .20 Hz), which can saturate the sensor’s response. Thus, the design and operation of infrasound sensors require special considerations to maximize inband sen- sitivity and dynamic range, and to reduce the influence of noise. Microphones sensitive to subaudible frequencies, mi- crobarometers, and differential pressure sensors have been adapted to measure the infrasonic range. These sensors use different mechanisms to convert pressure fluctuations into analog or digital signals. Proper cali- bration and characterization of infrasound sensors are required to ensure that this conversion provides an ac- curate representation of the infrasonic field (Kromer et al. 2007). This paper shows the implementation of an inexpensive low-power infrasonic sensor developed at the Infrasound Laboratory at the New Mexico In- stitute of Mining and Technology (Infra-NMT), which is based on a commercial micromachined differen- tial pressure transducer. The mechanisms used for signal transduction and noise filtering are explained, and the characterization and evaluation processes are described. 2. Infrasound sensors High-sensitivity barometers, used first to detect in- frasonic waves in the atmosphere (Bedard and Georges 2000), have been replaced by very specialized sensors intended to enhance the response in the infrasonic range. For example, the MB2000, MB2005 (CEA/ DAM 2009), and Chaparral 50 (Chaparral Physics 2011) [used in the Comprehensive Nuclear-Test-Ban Treaty (CTBT) International Monitoring System (IMS)] have noise floors that are 8 times lower than the mean atmospheric acoustic noise (Kromer and McDonald 2000) and are normally deployed along with atmo- spheric noise-reduction systems (Hedlin and Raspet 2003) to reduce the influence of the wind. This section provides an overview on the operation of various com- mercial infrasound sensors and compares their opera- tion to that of the Infra-NMT sensor introduced here. Operation of infrasound sensors Infrasonic sensors, or microbarometers, operate by measuring perturbations in the background atmospheric pressure (;100 kPa at sea level). Absolute and differ- ential pressure sensors can be used to detect infrasonic waves. An absolute pressure sensor would require high sensitivity and dynamic range to resolve infrasonic signals superimposed on atmospheric pressure. For example, an absolute pressure sensor that complies with IMS standards would cover the full extent of atmo- spheric pressure (;100 kPa at sea level) with a resolu- tion of at least 0.63-mPa rms (IMS sensor floor noise). This sensor would have a DR of 164 dB0.63 mPa rms, equivalent to 27 bits of resolution. This high resolu- tion in sensor output and digital conversion imply special requirements for datalogging capabilities and sensor operation, such as high-power consump- tion and low-noise electronic environments. Absolute microbarometers, such as the MB2000 or MB2005, provide barometric (absolute) and differential (high- pass-filtered version of the absolute channel) outputs with different sensitivities to facilitate digital conversion. Differential pressure sensors, on the other hand, measure the infrasound signal compared to a reference pressure (e.g., average atmospheric pressure). Again, using the IMS standards, a differential pressure sensor should resolve signals as small as 0.63-mPa rms and as large as 158-Pa rms. This sensor would require only an 18-bit digitizer to properly resolve the full range. Dif- ferential pressure sensors have two ports interconnected by a sensitive membrane or diaphragm that deforms with a differential pressure applied to the ports. Bellows (Jones and Forbes 1962), with silicon or metallic mem- branes, or a more complex system, such as fiber optics (Zumberge et al. 2003), have been used as sensing di- aphragms. Displacement sensors (Jiang et al. 2010), capacitors, or piezoresistive elements can be used to measure the diaphragm deflection (Zuckerwar et al. 2003) and convert that to an electrical signal. Condenser microphones (CMs), like those in phones and personal computers, use one of the plates of a ca- pacitor like a diaphragm; a pressure change over the plate deforms the diaphragm and changes the capaci- tance that can be detected with electronic circuits. CMs are mass produced to operate in audible frequencies (20–20 000 Hz) and can be used for infrasound de- tection if electronic conditioning is applied such that subaudible frequencies are amplified. However, the frequency response of these elements at low frequen- cies can vary from one element to the next, and re- peatability is not ensured. Nonetheless, the low cost of these elements made them popular as sensing elements for infrasonic sensors (Johnson et al. 2003: Ripepe et al. 2007). Piezoresistivity pressure transducers (PPTs), manu- factured using microelectromechanical systems (MEMS) technology, are extensively used in industrial and med- ical applications for low-pressure measurements (Barlian et al. 2009). Lower cost and higher performance in MEMS sensors have been driven by sustained research and development in MEMS technology (Ko 2007). In 1276 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 29 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 3 --- MEMS PPTs, materials that change resistance under applied stress are connected to micromachined silicon- based diaphragms. The deflection of the diaphragm and the corresponding change in resistance is measured us- ing a Wheatstone bridge (Bae et al. 2004). MEMS PPTs can sense static and dynamic pressures with high pre- cision. Because of the response of MEMS PPTs at very low frequencies, their use as a differential sensing ele- ment for an infrasound sensor depends on the ability to reject static pressure that can saturate the diaphragm and make them insensitive to pressures in the targeted frequency band. Pneumatic filters can provide a way to filter out the undesirable effect. The Infra-NMT sensor is based on a MEMS PPT transducer and uses a pneu- matic filter to reject saturation from very low frequency changes in the barometric pressure (,0.01 Hz). 3. Configuration of the Infra-NMT sensor The implementation of the infrasound sensor de- scribed here is based on the following three main com- ponents (see Fig. 1): 1) a micromachined pressure transducer (MEMS PPT), 2) an ultrastable voltage source, and 3) a mechanical filter based on a volume el- ement and a capillary tube. We used a 0.5 INCH-D-MV pressure transducer from All Sensors (All Sensor, Inc. 2011), which is linearly sensitive to static and dynamic pressures between 2124.5 and 124.5 Pa (differential 0.0-in. H2O pressure). This transducer has two ports (A and B) that are interconnected by a flexible silicon diaphragm (see Fig. 1b). A Wheatstone bridge is used to sense the deflection by measuring the changes in the piezoresistive elements attached to the diaphragm FIG. 1. Configuration of the Infra-NMT sensor. (a) Electronic scheme; (b) diagram of the mechanical filter and its elements and geometry; and (c) Infra-NMT enclosure, consisting of a PVC electric box with two outlets (one used for the analog output and power, and the other used to connect the pressure transducer to the atmosphere). SEPTEMBER 2012 M A R C I L L O E T A L . 1277 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 4 --- (Fig. 1a). When powered by a voltage source, the sen- sor’s output is proportional to supply voltage (ratio- metric). In this configuration, the sensitivity of the bridge is proportional to the voltage of the source. Therefore, the voltage input to the sensor’s transducer is important to keep stabilized and well regulated. To deliver a very stable and well-regulated voltage we used the voltage reference LT1021-7 (Linear Technology 2011). The LT1021-7 delivers the required voltage (7 V) with low noise (4 mV from peak to peak between 0.1 and 10 Hz), and consumes less than 2 mA; this reference voltage features also ultralow thermal drift (5 ppm 8C21), long-term stability [7 ppm (10000 h)21], and low line (1 ppm V21) and load regulation (12 ppm mA21). The LT1021-7 was chosen to reduce the influence of the Infra-NMT sensor to thermal changes and to ensure voltage stability in the long term. In addition, the voltage required for the operation of the regulator can be as low as 8 V, which allows the use of common 9- or 12-V batteries while maximizing the sensor’s sensitivity. In the Infra-NMT sensor, port A of the pressure transducer is open to the atmosphere and port B is connected to a closed volume chamber [also referred as backing volume, see Mutschlecner and Whitaker (1997)], with a capillary tube acting as a leak (Fig. 1b). The volume–capillary pair, connected to port B, acts like a low-pass filter for the atmospheric pressures outside our frequency band of interest. The difference in pres- sure between ports is a high-pass-filtered version of the atmosphere. The characteristics of the pneumatic filter are controlled by the geometry of the elements used in the filter. The next section provides an overview of the physics of this type of mechanical filter and a full de- scription of the characteristics of the filter used in the Infra-NMT sensor. Using a capillary tube and backing volume as a mechanical filter Mechanical, or pneumatic, filters based on passive acoustic elements (capillary tubes and chambers or volume elements) have been used along with differential pressure transducers to sense specific portions of the acoustic spectrum. For example, Richiardone (1993) used, and Mutschlecner and Whitaker (1997) described, mechanical filters to attenuate very low frequency at- mospheric events and high-frequency wind noise. Pneumatic filters are modeled by analogy with electric circuits, such that, depending on the geometry and size of acoustic elements, electronic counterparts can be identified. For example, sound moving through a cap- illary tube can be modeled as electrical current in a resistor. Thus, the capillary tube (assuming a small diameter compared to the length) can be characterized by its acoustic resistance, defined as (Keefe 1984) R 5 8h pr4l, (1) where h is the shear viscosity of the ambient air, and r and l the radius and length of the capillary, respectively. Ac- cording to Eq. (1), the longer and narrower the capillary tube, the higher its resistivity to airflow. Chambers or volume elements are able to temporarily ‘‘store’’ differ- ential pressure and behave like an electronic capacitor; these elements can be modeled by defining its acoustic capacitance as (Mutschlecner and Whitaker 1997) C 5 V gP 5 V rc2, (2) where V is the backing volume, c is the speed of sound, r is the air density, P is the average background pressure, and g 5 1.403 is the adiabatic gas constant, assuming adiabatic regime. The processes in the backing volume are not always in the adiabatic regime but can be in isothermal or transi- tional (from adiabatic to isothermal) regimes depending on the frequency range, air conditions, and geometry of the chamber. The transition frequency for the sensor’s backing volume (Mentink and Evers 2011), from more adiabatic to more isothermal behavior, is f 5 3.45 Hz, where f 5 h/pL2, h 5 k/rcp is the thermal diffusivity, k 5 0.025 W m21 K21 is the thermal conductivity, rcp 5 1.1 3 103 J m23 K21 is the heat capacity per unit volume, L 5 V/S is the characteristic size, V 5 1 cm3 is the internal volume, and S 5 7 cm2 is the internal surface. The value of g tends to 1.403 and 1 for the adiabatic (high frequencies) and isothermal limits (low frequencies). In the transition area, the value of g is replaced by (Richiardone 1993) g 5 Lg, (3) where jLj 5 f[A(g 2 1) 2 g]2 1[B(g 2 1)]2g1/2, arg(L) 5 (p/2) 1 tan21[A(g 2 1) 2 g=B(g 2 1)], A 1 iB 5 1 2 [2J1(z)=zJ0(z)] (assuming the chamber is a long cylinder), z 5 (22i) 1/2[rcpv/(2k)]1/2(V=S), v is angular frequency, and J0(z) and J1(z) are first-kind order 0 and 1 Bessel functions. Thus, g depends of the frequency and the filter geometry and has values between 1 and 1.403. Figure 2b shows how g changes in the frequency range of interest for the Infra-NMT sensor. The parameters of acoustic resistance (R) and ca- pacitance (C; with the correction for the frequency de- pendent coefficient g) are used to model the response to this acoustic low-pass filter using the electronic anal- ogy of an RC filter connected in series, 1278 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 29 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 5 --- fcutoff 5 1 2pRC 5 rc2 16h r4 Vl   5 gP 16h r4 Vl   5 gLP 16h r4 Vl   . (4) This acoustical low-pass filter is connected to one of the ports of the differential pressure sensor and the other is open to the atmosphere. The differential pres- sure across the diaphragm [Dp(t) in time or DP(v) in frequency] is the difference in pressure between un- filtered and filtered ports, Dp(t) 5 p(t) 2 hlp(t)*p(t), (5) where the asterisk (*) is the convolution operator, p(t) and P(v) are the atmospheric pressure in time and fre- quency, and hlp(t) and Hlp(v) 5 1/(1 1 ivRC) is the impulse and frequency response of an acoustic single-pole low-pass filter. Using the convolution theorem, the frequency counterpart of Eq. (5) can be written as DP(v) 5 P(v)[1 2 Hlp(v)] or DP(v) 5 P(v)  1 2 1 1 1 ivRC  . (6) Equation (6) can be reorganized as DP(v) 5 P(v)  ivRC 1 1 ivRC  , where the term Hhp(v) 5 ivRC/(1 1 ivRC) is the fre- quency response of a single-pole high-pass filter with corner frequency controlled by the product RC; the correspond- ing impulse response for this filter is hhp(t). Equations (5) and (6) can be rewritten as Dp(t) 5 hhp(t)*p(t) and FIG. 2. (a) Theoretical and experimental frequency response for the Infra-NMT sensor with standard and test filters. The theoretical frequency responses for a standard Infra-NMT (blue) and test filters (magenta) assuming adiabatic (dashed lines) and transitional (solid lines) regimes, respectively. The frequency re- sponses of two Infra-NMT sensors (green and cyan lines) are shown, as is the frequency response of the test filter sensor (black line with circles). The models assuming high (light gray; 15% in capillary tube length and backing volume and 5% capillary tube radius) and low (dark gray; 10% in capillary tube length and backing volume and 5% capillary tube radius) tolerances are represented, respectively. These tolerance areas are centered at the transitional and adiabatic responses for the standard Infra-NMT and test sensor, re- spectively. The transition frequency from adiabatic to isothermal regimes (plain black line) is shown. (b) g and the correction factor L. In the frequency range of the Infra-NMT, g is replaced by g 5 gL. SEPTEMBER 2012 M A R C I L L O E T A L . 1279 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 6 --- DP(v) 5 P(v)Hhp(v), respectively. Thus, the entire sys- tem behaves as a single-pole high-pass filter. Table 1 shows the parameters used and the corresponding corner fre- quencies for mechanical filters in different infrasound sensors that are used for sensor evaluation of the Infra- NMT. In these calculations, the molecular shear viscosity is 1.8013 kg m21 s21 and the atmospheric pressure is equal to 81 kPa; both are considered constants. Figure 2a shows the theoretical and experimental frequency response of the filters described in Table 1 for the Infra-NMT sensor implementation. The solid and dashed blue lines are the theoretical frequency re- sponses for a standard Infra-NMT filter, assuming adi- abatic and transitional regimes, respectively. The green and cyan lines show the frequency responses of two Infra-NMT sensors. The light and dark gray areas are centered at the transitional response (for the standard Infra-NMT) and represent the model assuming high and low tolerances, respectively, for the geometrical para- meters of the mechanical filter. These tolerances are introduced to account for variations resulting from the manual construction of the mechanical filter. Sensors 1 and 2 agree better with the transitional than adiabatic model. However, sensor 2 still lies outside of the higher- tolerance region; this deviation can be related to a par- ticularly long capillary tube and/or a big backing volume. Also, a nonstandard Infra-NMT mechanical filter (with a wider capillary tube) was tested (black line with cir- cles). The response for this test filter is closer to the adiabatic (tolerance areas for this sensor are centered at the adiabatic model) than transitional response. How- ever, this response is also outside of the high-tolerance area, and we suspect a particularly short capillary tube and/or backing volume. The diameter of the capillary tube is the most stable of the geometrical parameters of the filter and the tolerance was kept at 5%. 4. Sensor characterization and evaluation: Double- reference technique to determine the Infra-NMT frequency response The Infra-NMT infrasound sensor is configured to re- spond to a wide frequency range (i.e., 0.01–40 Hz) with a flat response and sensitivity of 46.85 mV Pa21 (using manufacturer’s specifications). This sensitivity is de- termined by scaling the nominal span, that is, 10 mV for 124.5 Pa (0.5 in. of H2O) at 12 V, to the new voltage reference (7 V). The response time of the pressure sensor to get 90% of any applied pressure is less than 100 mS; this ensures a high-frequency corner of several hundreds to a few thousands of hertz, which is well above the infrasound band. The low-frequency corner of 0.01 is controlled by geometry of the elements of the mechanical filter, where the capillary tube has a length of 1.1 cm and a diameter of 25 mm, and the volume of the chamber is 1 cm3. The total current consumption (3 mA) was determined by adding the consumption of the voltage reference (1.8 mA) and the consumption of the transducer (1.2 mA). To evaluate the performance of this sensor, the in- frasonic testing chamber at Sandia National Laboratory (SNL) Facility for Acceptance, Calibration, and Testing (FACT) was used. Figure 3 shows a diagram of the calibration system, where a Smart24 digitizer (Geotech Instruments 2011) provides up to six channels for analog- to-digital conversion. The active element in this testing setup is based on a mechanical piston (APS300; APS TABLE 1. Parameters used in different sensors in acoustic filters. Equation (3) was used for the determination of the cutoff frequency using h 5 1.8013 and atmospheric pressure of 81 kPa. Sensor Capillary tube Backing volume Cutoff frequency Length (m) Radius (mm) (1026 m3) Model (Hz) Expt (Hz) Mutschlecner and Whitaker (1997) 0.5 200 200 0.0062 Richiardone (1993) 1.012 254 10 690 0.0001 Infra-NMT (sensor 1) 0.01 25 1 0.01 0.004 Infra-NMT (sensor 2) 0.01 25 1 0.01 0.01 Infra-NMT (test filter) 0.01 75 1 1.22 2.5 FIG. 3. Infrasound sensor testing chamber at the SNL FACT site. The chamber is located 5 m below ground to prevent that diurnal temperature variations and wind noise affect the evaluation. 1280 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 29 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 7 --- Dynamics 2011) that is connected to one of the digital- to-analog channels of the Smart24 digitizer. The cham- ber has a nominal volume of 50 L, which was reduced to 30 L to allow the generation of high-pressure signals (e.g., signals with full-scale amplitudes in the 10–150-Pa range). The infrasound test chamber is installed in an underground enclosure, approximately 5-m under- ground, which reduces the influence of diurnal tem- perature variation and wind noise on the component evaluation. Standard infrasound sensor evaluation, following comparison calibration procedures, uses one reference (e.g., MB2000) to evaluate the new device under test. However, because of the broad frequency response of the Infra-NMT sensor design (i.e., 0.01–40 Hz), which in- cludes relatively high frequencies, a ‘‘double reference’’ sensor scheme was setup using an MB2000 micro- barometer and a calibrated GRAS 42AE microphone (GRAS Sound and Vibration 2008). The MB2000 mi- crobarometer features flat response (within 3dB) for frequencies between 0.01 and 8 Hz and noise levels less than 2 mPa between 0.02 and 4 Hz. The GRAS 42AE microphone features a flat response (within 3 dB) from 1 to 100 Hz. By overlapping the flat responses of these two well-calibrated references we are able to comment of the full range of the response of the Infra-NMT sensor. Two types of tests were performed on a group of seven Infra-NMT sensors, using the standard mechan- ical filter. The testing tools at the SNL FACT site are fully described by Kromer et al. (2007) and include 1) Infrasound Sensor Piston-Phone Linearity Verifica- tion (IS-PLV), which uses sinusoid pressure waves of increasing amplitude at a single frequency to comment on the sensors linearity of voltage output versus pres- sure; and 2) Infrasound Sensor Frequency/Amplitude/ Phase Response Verification (IS-FAPR), which uses a broadband white noise pressure signal to verify the amplitude and phase response of the sensor design across the frequencies of interest (i.e., 0.01–40 Hz). The IS-PLV, for this evaluation, used a single tone of 1 Hz and amplitudes of 0.05, 0.2, 0.4, 2, 4, 23, 50, 100, and 121 Pa on the seven Infra-NMT sensors. The sensors showed linear responses in the tested amplitudes (see Fig. 4a), and we expect the sensors to be linear in the manufacturer’s full-scale linear range (up to 124 Pa). Results from the IS-PLV test show an average sen- sitivity of 45.13 mV P21, with a standard deviation of 0.23. Figure 4b shows the calculated sensitivity at the different tested amplitudes by using the estimated noise FIG. 4. IS-PLV results and analysis: (a) voltage response of the Infra-NMT sensor vs applied pressure, and the linear regression of sensor output voltage (Y) vs input pressure (X) and the coefficient of determination; (b) sensitivity of the Infra-NMT. The sensitivity of the Infra-NMT is determined by averaging the sensitivity of all sensors at the different amplitudes. The con- fidence intervals are determined by using the noise’s rms, determined for each input pressure. SEPTEMBER 2012 M A R C I L L O E T A L . 1281 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 8 --- during the IS-PLV test for individual sensors; uncertainty intervals are estimated for sensitivity. For low-pressure amplitudes, these intervals are big because the sensor noise and input signal have comparable levels (low signal- to-noise ratio); for high amplitudes, the intervals increase due to distortions in the pressure generator and/or re- sponse of the chamber at higher pressures. For the IS-FAPR test, we used a normally distributed white noise signal with a zero-peak amplitude of 23 Pa and 1000-Hz bandwidth. A two–channel coherence tech- nique (Merchant and Hart 2011) was applied to two Infra-NMT sensors to determine the noise spectrum and confidence intervals (Fig. 5). Using this frequency spectrum, noises in specific bands were determined as follows: 1) 2.02-mPa rms between 0.5 and 2 Hz, 2) 5.47-mPa rms between 0.1 and 20 Hz, and 3) 5.62-mPa rms between 0.05 and 20 Hz. These intervals were se- lected to facilitate comparison with other sensors. Table 2 summarizes the complete set of characteristics of the Infra-NMT sensor determined by the evaluation process. The cost of the elements for the implementation of an Infra-NMT single unit is several hundred dollars. 5. Conclusions We have packaged and calibrated a low-cost infra- sound sensor with low noise (2.2-mPa rms at 0.5–2 Hz); high dynamic range (up to 95 dB for the 0.5–2-Hz band); and sensitivity of 45.13 mV Pa21, which is appropriate for infrasonic studies that may encounter high levels of background infrasound noisy and/or near-source en- vironments. Its low-cost, low-power consumption, and small size (cf. to other infrasound sensors with higher sensitivity and lower noise) favor its use in network and array configuration requiring large quantities of sensors, in near-source locations (where pressure signals are in- tense), and/or in higher noise environments. The Infra- NMT has been used successfully in many environments and applications [e.g., volcano (Goto and Johnson 2011), atmospheric (Marcillo and Johnson 2010), and thunder (Arechiga et al. 2010) studies]. A simple model FIG. 5. Self-noise spectrum of the Infra-NMT sensor. This self-noise response is shown of the Infra-NMT sensor, the acoustic low-noise model (ALNM) by Bowman et al. (2005), and an example signal from a 500-lb ammonium nitrate/fuel oil (ANFO) explosion recorded at 5.5 km. TABLE 2. Summary of the characteristics of the Infra-NMT sensors. Parameter Value Units Sensitivity 45.13 6 0.23 mV Pa21 Nominal linear range 6124.5 Pa Noise mPa rms 0.5–2 Hz 2.02 0.1–20 Hz 5.47 0.05–20 Hz 5.62 Power consumption (using a 9-V source) 27 mW Sensor voltage range (SVR) 8 # SVR # 40 V Current (using a 9-V source) 3 mA 1282 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 29 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 9 --- describing the mechanical filter was tested by comparing theoretical and experimental responses of different filter implementations. We suspect that deviations from the ideal response are related to the construction of the mechanical filter. However, assuming a tolerance up to 15% in the parameters controlled by the construction (capillary tube length and backing volume), the Infra- NMT provides a flat response (within 3 dB) throughout the infrasonic range above the specified corner fre- quency. The implementation of the double-reference scheme in the testing chamber at the SNL FACT site provided an effective tool to evaluate the wide fre- quency response of the Infra-NMT sensors. Acknowledgments. This work was partially supported through funding from NSF EAR 0838562 and NSF EAR 0738802. REFERENCES All Sensor, Inc., 2011: Millivolt output pressure sensors. Datasheet Rev. C, 4 pp. [Available online at http://www.allsensors.com/ datasheets/DS-0091-Rev-C1.pdf.] Antier, K., A. Le Pichon, S. Vergniolle, C. Zielinski, and M. Lardy, 2007: Multiyear validation of the NRL-G2S wind fields using infrasound from Yasur. J. Geophys. Res., 112, D23110, doi:10.1029/2007JD008462. APS Dynamics, 2011: APS 300 portable shaker-amplifier. Datasheet, Rev. April 2011, 2 pp. [Available online at http:// www.apsdynamics.com/images/stories/Prospekte/APS_Shaker/ APS_300/APS_300_Data_Sheet_en.pdf.] Arechiga, R. O., J. B. Johnson, H. E. Edens, R. J. Thomas, and W. Rison, 2010: Acoustic localization of triggered lightning. J. Geophys. Res., 116, D09103, doi:10.1029/2010JD015248. Arrowsmith, S. J., J. B. Johnson, D. P. Drob, and M. A. H. Hedlin, 2010: The seismoacoustic wavefield: A new paradigm in study- ing geophysical phenomena. Rev. Geophys., 48, RG4003, doi:10.1029/2010RG000335. Bae, B., B. R. Flachsbart, K. Park, and M. A. Shannon, 2004: Design optimization of a piezoresistive pressure sensor con- sidering the output signal-to-noise ratio. J. Micromech. Mi- croeng., 14, 1597, doi:10.1088/0960-1317/14/12/001. Barlian, A., W. Park, J. Mallon Jr., A. Rastegar, and B. Pruitt, 2009: Review: Semiconductor piezoresistance for microsystems. Proc. IEEE, 97, 513–552. Bedard, A. J., and T. M. Georges, 2000: Atmospheric infrasound. Phys. Today, 53, 32. Bowman, J. R., G. E. Baker, and M. Bahavar, 2005: Ambient in- frasound noise. Geophys. Res. Lett., 32, L09803, doi:10.1029/ 2005GL022486. CEA/DAM, 2009: MB2000 and MB2005 microbarometers. [Available online at http://www-dase.cea.fr/public/dossiers_ thematiques/microbarometres/description_en.html.] Chaparral Physics, 2011: Operation manual for the Model 50A infrasound sensor. Operation manual, Rev. May 2010, 17 pp. [Available online at http://www.gi.alaska.edu/files/Model %2050A%20Manual%20beta.pdf.] Geotech Instruments, 2011: The Smart Series. Datasheet Rev. August 2010. [Available online at http://www.geoinstr.com/ ds-smart24.pdf.] Goto, A., and J. B. Johnson, 2011: Monotonic infrasound and Helmholtz resonance at Volcan Villarrica (Chile). Geophys. Res. Lett., 38, L06301, doi:10.1029/2011GL046858. GRAS Sound and Vibration, 2008: Low-frequency calibrator type 42AE. Datasheet, Rev. March 2008. [Available online at http://www.grasinfo.dk/documents/pd_42AE_ver_19_03_ 2008.PDF/.] Hedlin, M. A. H., and R. Raspet, 2003: Infrasonic wind-noise re- duction by barriers and spatial filters. J. Acoust. Soc. Amer., 114, 1379–1386. Jiang, B., K. Yang, and J. Wang, 2010: Design of the opti- cal fiber MEMS infrasound sensor. Proc. 2010 Int. Conf. Measuring Technology and Mechatronics Automation (ICMTMA), Changsha, China, IEEE Computer Society, 1049–1051. Johnson, J. B., R. C. Aster, M. C. Ruiz, S. D. Malone, P. J. McChesney, J. M. Lees, and P. R. Kyle, 2003: In- terpretation and utility of infrasonic records from erupting volcanoes. J. Volcanol. Geotherm. Res., 121, 15–63. Jones, R., and S. Forbes, 1962: A microbarograph. J. Sci. Instrum., 39, 420–427. Keefe, D. H., 1984: Acoustical wave propagation in cylindrical ducts: Transmission line parameter approximations for iso- thermal and nonisothermal boundary conditions. J. Acoust. Soc. Amer., 75, 58–62. Ko, W. H., 2007: Trends and frontiers of MEMS. Sens. Actuators A, 136, 62–67. Kromer, R. P., and T. S. McDonald, 2000: Infrasound sensor models and evaluation. Proc. 22nd Annual DoD/DOE Seismic Research Symp.: Planning for Verification of and Compliance with the Comprehensive Nuclear-Test-Ban Treaty (CTBT), New Orleans, LA, Department of Defense and Department of Energy. [Available online at http://handle.dtic.mil/100.2/ ADA523399.] ——, D. M. Hart, and J. M. Harris, 2007: Test definitions for the evaluation of infrasound sensors. Sandia National Laborato- ries Tech. Rep. SAND2007-5038, 11 pp. [Available online at http://prod.sandia.gov/techlib/access-control.cgi/2007/075038. pdf.] Le Pichon, A., E. Blanc, and D. Drob, 2005: Probing high-altitude winds using infrasound. J. Geophys. Res., 110, D20104, doi:10.1029/2005JD006020. Linear Technology, 2011: LT1021—Precision Reference. Da- tasheet, Rev. C. [Available online at http://cds.linear.com/ docs/Datasheet/1021fc.pdf.] Marcillo, O., and J. B. Johnson, 2010: Tracking near-surface at- mospheric conditions using an infrasound network. J. Acoust. Soc. Amer., 128, EL14–EL19. Mentink, J. H., and L. G. Evers, 2011: Frequency response and design parameters for differential microbarometers. J. Acoust. Soc. Amer., 130, 33–41. Merchant, J., and D. M. Hart, 2011: Component evalua- tion testing and analysis algorithms. Sandia National Lab- oratory SAND2011-8265, 136 pp. [Available online at http://prod.sandia.gov/techlib/access-control.cgi/2011/118265. pdf.] Mutschlecner, J. P., and R. W. Whitaker, 1997: The design and operation of infrasonic microphones. Los Alamos National Laboratory Tech. Rep. LA-13257, 27 pp. [Available online SEPTEMBER 2012 M A R C I L L O E T A L . 1283 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC --- Page 10 --- at http://www.osti.gov/bridge/servlets/purl/481856-QT6CGV/ webviewable/481856.pdf.] Richiardone, R., 1993: The transfer function of a differential mi- crobarometer. J. Atmos. Oceanic Technol., 10, 624–628. Ripepe, M., E. Marchetti, and G. Ulivieri, 2007: Infrasonic moni- toring at Stromboli Volcano during the 2003 effusive eruption: Insights on the explosive and degassing process of an open conduit system. J. Geophys. Res., 112, B09207, doi:10.1029/ 2006JB004613. Vergniolle, S., M. Boichu, and J. Caplan-Auerbach, 2004: Acoustic measurements of the 1999 basaltic eruption of Shishaldin Volcano, Alaska: 1. Origin of Strombolian activity. J. Volca- nol. Geotherm. Res., 137, 109–134. Zuckerwar, A. J., T. R. Kuhn, and R. M. Serbyn, 2003: Back- ground noise in piezoresistive, electret condenser, and ce- ramic microphones. J. Acoust. Soc. Amer., 113, 3179– 3187. Zumberge, M. A., J. Berger, M. A. H. Hedlin, E. Husmann, S. Nooner, R. Hilt, and R. Widmer-Schnidrig, 2003: An optical fiber infrasound sensor: A new lower limit on atmospheric pressure noise between 1 and 10 Hz. J. Acoust. Soc. Amer., 113, 2474. 1284 J O U R N A L O F A T M O S P H E R I C A N D O C E A N I C T E C H N O L O G Y VOLUME 29 Unauthenticated | Downloaded 09/17/26 05:50 PM UTC